Valco Instruments Co. Inc.
Pulsed Discharge Detector
Models D-4-I-TQ-R
and D-4-I-TQI-R
Instruction Manual
Rev 1/15
Valco Instruments Co. Inc.
800 · 367· 8424 sales
713
· 688· 9345 tech
713
· 688· 8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
· 41 · 925· 6200 phone
Int + 41
· 41 · 925· 6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact::
Seitenansicht 0
1 2 3 4 5 6 ... 21 22

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Valco Instruments Co. Inc.Pulsed Discharge Detector Models D-4-I-TQ-R and D-4-I-TQI-R Instruction Manual Rev 1/15 Valco Instruments Co. Inc.800 · 3

Seite 2

6InstallationGeneral Precautions • Do not use plastic/polymer or copper tubes for gas handling andinterconnectons. Use only stainless steel tubing w

Seite 3 - Table of Contents

7InstallationGAS CHROMATOGRAPHTEE(ZT1)HELIUMPURIFIERDISCHARGE GAS(99.999% He)TGA-R-30F60PRESTRICTOR(30 mL/min minimum)VENTCOLUMNINLETDISCHARGE GASINLE

Seite 4

8Installation7. Open the cylinder valve to pressurize the regulator once again. Close thevalve and observe the needle of the high pressure gauge for

Seite 5 - Introduction

9Detector Electrical ConnectionsCAUTION: Do not use a wrench to tighten the SMCconnectors on the bias and electrometer cables.Connections should be fi

Seite 6 - Safety Notes and Information

10Initial Power-UpCAUTION: Always make sure that discharge gas isflowing before heating and powering up the detector.1. Before installing the column,

Seite 7

11InstallationCapillary Column ConnectionIf the capillary column adapter is installed in the column inlet:1. Make a mark on the column 14.5 cm from th

Seite 8 - The discharge gas

123. Set aside the 1/8" wrench and completely remove the adapter from thecolumn inlet.4. Screw the packed column adapter into the column inlet by

Seite 9 - Gas Specifications

13Mode Selection and SetupHelium Ionization ModeIf the instructions at the top of page 11 were properly executed, the columnshould already be properly

Seite 10 - Installation

14Mode Selection and SetupAr-PDPIDChanging the discharge gas from helium to a mixture of 2% argon inhelium changes the photon energy level from the 17

Seite 11 - Gas Connections

15TroubleshootingHigh Background CurrentLow SensitivityPurple/bluePinkDisconnect the columnand cap the column inletBaseline current is still high(>

Seite 12

This page intentionally left blank for printing purposes

Seite 13

16High Noise LevelIf the noise lievel is high:1. See if it improves with the GC fan turned off.2. Check the column insertion position (14.5 cm for cap

Seite 14 - Initial Power-Up

17WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Seite 15 - Packed Column Connection

18Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up,page 10), the standing

Seite 16 - Exercise

Table of ContentsIntroduction Description and Operating Principles ...1 Safety Notes an

Seite 17 - Mode Selection and Setup

This page intentionally left blank for printing purposes

Seite 18

1IntroductionDescription and Operating PrincipleThe PD-D-4-I-TQI-R and PD-D-4-I-TQ-R are non-radioactive pulsed dischargeionization detectors (PDIDs)

Seite 19 - Troubleshooting

2Helium Ionization (PDHID) ModeThe PDHID is essentially non-destructive (0.01 - 0.1% ionization) and highlysensitive. The response to organic compoun

Seite 20 - High Noise Level

3Components of the Detector SystemComponents of the detector systems are listed in the tables below. Check the package contents to verify that everyt

Seite 21 - Warranty

4IntroductionSystem RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high puri

Seite 22 - Detector Performance Log