Vici 500 Dynacalibrator Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messgeräte Vici 500 Dynacalibrator herunter. VICI 500 Dynacalibrator User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 37
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
Valco Instruments Co. Inc.
9/13
Dynacalibrator
Model 500
Instruction Manual
Valco Instruments Co. Inc.
800 · 367· 8424 sales
713
· 688· 9345 tech
713
· 688· 8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
· 41 · 925· 6200 phone
Int + 41
· 41 · 925· 6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact::
®
Manufactured by:
Valco Instruments Co. Inc.
8300 Waterbury
Houston, TX 77055
Seitenansicht 0
1 2 3 4 5 6 ... 36 37

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Valco Instruments Co. Inc. 9/13 Dynacalibrator Model 500 Instruction ManualValco Instruments Co. Inc.800 · 367· 8424 sales 713 · 688· 9345 tech 713

Seite 2 - Symbols used in this document

6Getting StartedEvery Dynacalibrator is completely calibrated, thoroughly tested and inspected, and carefully packed prior to shipment. The carrier h

Seite 3 - Table of Contents

7Concealed DamageConcealed damage is damage which is not apparent until after the equipment has been unpacked and examined or tested. In the event tha

Seite 4 - Table of Contents (continued)

8Unit LocationIn general, the Dynacalibrator should be as close as possible to the analyzer to be calibrated, as long as:• the ambient temperature is

Seite 5 - Introduction

9Getting StartedRear Panel ConnectionsRemove all shipping caps and shipping plugs from the rear panel bulkhead fittings. All these hardware items sho

Seite 6

10Carrier/Dilution InletsAll 500 Series Dynacalibrators are configured with either a single inlet (500-0-***) or a dual inlet (500-1-***) and therefore

Seite 7

11Front Panel Controls and IndicatorsThe section entitled “Advanced Theory of Operation” beginning on page 24 includes more detailed information on t

Seite 8 - Special Customer Options

12temperature can be set manually through the controls on the front panel or remotely through RS-232 communication. (Refer to the chapter entitled “

Seite 9 - Specifications

13CalculationsConcentration of the permeant compound in the span outlet stream is inversely proportional to the carrier flow rate through the chamber.

Seite 10 - Getting Started

14Initial Power-UpConnections1. Plug the supplied power cord into the rear mounted power supply. Ensure that the power supply output cable is plugge

Seite 11 - Concealed Damage

15Setting the Chamber Temperature ManuallyTo read the current set point: press and hold the button. To read the chassis environment temperature: p

Seite 12 - Unit Location

AttentionRadiant heat warningEntanglement warningElectrical warningSymbols used in this document

Seite 13 - Rear Panel Connections

16Setting the Chamber Temperature via Serial Port (RS-232)Refer to the section titled “Serial Port Communications” on the next page.Installing the Per

Seite 14 - Span and Zero Outlets

17Serial Port CommunicationThe Dynacalibrator Model 500 can be monitored and controlled remotely by means of a serial port connection and a terminal e

Seite 15 - 

182. Click “Cancel”. Since no modem is involved, this information is not necessary. The following screen will appear:3. Click “OK”, which reveals

Seite 16

196. Select the appropriate COM port and click “OK” to bring up this screen:7. Set the port parameters as indicated above, and click “OK”.8. Now we

Seite 17 - Calculations

209. On the dyna500 Properties screen (above), click “Settings” to bring up the screen below. Click “Settings”Click “ASCII Setup...”Serial Port Commu

Seite 18 - Initial Power-Up

2110. Click “ASCII Setup” and set the parameters as indicated below. Click “OK”.11. When the screen returns to Dyna500 HyperTerminal (page 19, botto

Seite 19

22Analyzer CalibrationAfter the power-up procedure has been completed and the chamber temp-erature and permeation device have reached equilibrium, the

Seite 20

23Shutdown ProcedureDynacalibrators are designed to operate continuously; however, if it is necessary to turn the unit off, use the following procedur

Seite 21 - Serial Port Communication

24Figure 5: Basic block diagramAdva

Seite 22

25Plumbing ConfigurationsThere are two plumbing configurations available for Model 500 Dynacalibrators. The plumbing configuration is defined by the first

Seite 23

Table of ContentsIntroduction General Description...

Seite 24

26Figure 8: Model 500 electrical schematic Electrical DescriptionThe basic wiring of the Model 500 is shown schematically in Figure 8. Dynacalibrato

Seite 25 - Entering Commands

27MaintenanceSince Dynacalibrators are designed for and generally used in applications that require continuous service, a planned routine maintenance

Seite 26 - Analyzer Calibration

289. Inspect all assembly mountings for loose hardware, and tighten as necessary.10. Rinse the fan screen and fan filter in clean water and blow dry

Seite 27 - Shutdown Procedure

29MaintenanceChamber Temperature CalibrationPrior to shipment, the permeation chamber digital temperature control is calibrated against the temperatu

Seite 28 - Advanced Theory of Operation

30TroubleshootingIf you intend to perform detailed troubleshooting on your Dynacalibrator, please first read the section entitled “Advanced Theory of

Seite 29 - Plumbing Configurations

315. Repeat steps 3 and 4 for each potential leak point in the suspect loop.6. Tighten or remake any connection or joint that has been found to be l

Seite 30 - Electrical Description

32Model Number Breakdown  

Seite 31 - Maintenance

33WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyer may also have other rights that vary from state to state. For a per

Seite 32 - Carrier Flow Rate Verification

Table of Contents (continued)Electrical Description ...

Seite 33

1IntroductionGeneral DescriptionVICI Metronics Dynacalibrators® use Dynacal® permeation devices to generate the precise gas concentrations necessary f

Seite 34 - Troubleshooting

2

Seite 35 - Component Isolation Method

3IntroductionReferring again to Figure 1, note that the dilution stream passes through scrubber #3 and feeds into differential pressure regulator DPR

Seite 36 - Model Number Breakdown

4The Model 500 has four modes of operation – ZERO, SPAN 1, SPAN 2, and Both SPAN 1 and SPAN 2. When the calibrator mode is changed to ZERO, SPAN 1, o

Seite 37 - Warranty

5IntroductionSpecifications

Verwandte Modelle: Dynacalibrator 500

Kommentare zu diesen Handbüchern

Keine Kommentare